Skip to Content
Reading an IOL Defocus Curve: What 'Extended Depth of Focus' Means on the Graph (2026 Guide)
Every EDOF claim sits on one graph. How to read the axes, the 0.2 logMAR and 0.5 D EDOF tests, and why chart type, curve fitting and pupil size change what the curve shows.
Reading an IOL Defocus Curve: What EDOF Means on the Graph | Agaaz Ophthalmics

Refractive Solutions · Surgeon Series

Every EDOF claim sits on
one graph. Here is how
to read it.

A defocus curve shows how vision changes as focus is shifted away from distance. It is the graph behind every “extended depth of focus” brochure. The shape depends on the chart, the threshold, the pupil and the fitting method, so the same lens can look different on two different graphs. This guide covers the axes, the EDOF tests, and what the pooled trials do and do not show.

0.2 logMARthe acuity threshold
in the EDOF test
0.5 Dminimum depth-of-field
gain over a monofocal
1 D = 1 mdefocus converts to
viewing distance

Section 01 — Why One Number Misleads

A single “depth of focus”
figure hides the shape.

Brochures often reduce a lens to one number: a depth of focus in dioptres, or a single acuity value at one distance. A single number can be true and still tell a surgeon very little. Two lenses can share the same figure and have very different ranges of good vision.

Buckhurst and colleagues tested this directly with multifocal lenses. They found that single-value depth-of-focus metrics could not separate the designs, and proposed an area-of-focus measure that tracked what patients perceived at intermediate and near distances. The lesson for reading any brochure is simple: a curve carries information that one number discards.

The core idea in one paragraph

A defocus curve plots visual acuity against the amount of defocus. The x-axis is how far focus is shifted from distance, and the y-axis is acuity, usually in logMAR, where lower is better. Read the curve as a range: how wide a band of defocus keeps acuity above the threshold you care about. Every number in the brochure is a point on that curve, chosen by someone.

The x-axis
Defocus, in dioptres, from distance focus toward nearer focus. Check whether the lens is tested at a fixed pupil and whether the axis is labelled in dioptres or in viewing distance.
The y-axis
Visual acuity, usually logMAR. Lower values are better. Confirm the chart used to measure it, because the chart changes the value.
The threshold
The acuity line that defines “in focus”. A curve that looks wide at a lenient threshold can look narrow at a strict one.
The range
The width of the band of defocus that stays above the threshold. This is the quantity EDOF claims are really about.

Section 02 — Interactive

Dioptres are distances
in disguise.

Defocus is expressed in dioptres, but patients think in metres. A dioptre is the reciprocal of a distance in metres, so the conversion is simple arithmetic. Use the slider to see which viewing distance a given amount of defocus corresponds to.

Interactive: defocus to viewing distance
Pure arithmetic, not a measurement: a defocus of D dioptres corresponds to an object at 1 / D metres. Nothing on this widget describes any lens.
1.00 D
DefocusViewing distance equivalentTypical brochure use
0.5 D2 mGain threshold in the EDOF test
1.0 D1 mIntermediate range (computer distance sits near here)
1.5 Dabout 67 cmBetween intermediate and near
2.0 D50 cmNear reading distance
2.5 D40 cmClose near

Conversions are exact to rounding. The brochure column describes common usage only.

Section 03 — What Counts as EDOF

The test behind
the label.

“EDOF” is not a regulated marketing term with one number attached. Researchers classify a lens against a monofocal reference using defined tests, and those tests are what make the label comparable from one study to the next.

A scoping review by Fernández and colleagues (2023), which set enhanced monofocal lenses against conventional monofocal and EDOF lenses, summarises an operational EDOF test built on two conditions: a gain of at least 0.5 D in depth of field at 0.2 logMAR, and intermediate visual acuity of 0.2 logMAR or better, both compared with a monofocal.

Why the threshold matters. The 0.2 logMAR line is a convention. If a brochure reports a range at 0.3 logMAR, or does not state the threshold at all, the range is not directly comparable to a claim made at 0.2. Always ask which threshold the curve was drawn against.

Section 04 — The Same Lens, Different Curves

Three things change
the curve without
changing the lens.

Two studies of the same optic can produce different curves. Three methodological choices explain most of the disagreement, and each one is visible if you know where to look.

1. The chart
Tana-Rivero and colleagues compared recognition and resolution charts for defocus curve determination in trifocal lenses. Chart type alters the measured acuity at the same defocus, so curves from different studies are not directly interchangeable.
2. The curve fit
Law and colleagues found that the fitting method changes the result. Some polynomial fits over- or under-fit the curves of EDOF and monofocal lenses, while spline-based approaches were more consistent across lenses. A fitted curve is a model of the data, not the data.
3. Steps and range
The defocus steps tested and the total range covered decide how finely the curve is sampled. A curve sampled coarsely can hide a dip, and a curve that stops early can look wider than it is.
Eyes, binocularity and n
Monocular and binocular curves differ, and a curve from a small group of eyes carries wide uncertainty. Check the number of eyes and whether testing was monocular.
Reading a brochure, step by step
What a single graph should disclose

Chart. Threshold. Pupil diameter. Number of eyes. Monocular or binocular. Defocus step and total range. Fitting method. If a graph is missing any of these, treat its width as unverified rather than as a fact about the lens.

Section 05 — Pupil and Cornea

The pupil is part of
the optical system.

A defocus curve is not a property of the lens alone. The eye's own spherical aberration and the pupil diameter decide how light from different foci combines on the retina.

Chae and colleagues evaluated two extended-depth-of-focus lenses on the bench, using model corneas. Their through-focus measurements showed that the pattern depended on pupil size and corneal spherical aberration: at some simulated pupil diameters the through-focus response looked bifocal-like, and at a smaller pupil the foci merged into a single broad zone. This is a bench study of two lenses, and it does not measure patient outcomes, but it shows why a curve measured at one pupil size cannot be assumed for another.

Kozhaya and colleagues used an adaptive-optics simulator to vary spherical aberration in pseudophakic eyes. Their finding, that sensitivity to spherical aberration differs between monofocal and EDOF designs, is a simulation result rather than an implanted-lens outcome. Tabernero and colleagues found that depth of focus changed with induced negative spherical aberration, and that the response varied widely between people, with some gaining depth of focus and others not.

The practical point. Ask whether a curve was measured at a pupil size that resembles your patient's. Night-time pupils are larger, and the same lens can present a different through-focus pattern at that size.

Section 06 — Schematic

What the curve shapes
look like, in outline.

Select a lens type to see the general shape its curve is usually described with. These outlines are illustrative sketches to show what the terms mean. They contain no measured values, and they are not drawn from any particular lens.

Schematic: monofocal, EDOF and trifocal outlines
Schematic. Shape only. No axis values are shown and none are implied. Real curves must be read from the manufacturer's or study's own data with its chart, threshold and pupil stated.
Distance end of the axis is at the left; nearer focus to the right

The three outlines capture the design intent. A monofocal curve is a narrow peak around distance. An EDOF curve is a broader, flatter shoulder that keeps acuity usable across intermediate distances. A trifocal curve shows peaks for distance and near, with a dip between them. The schematic does not tell you how deep any real shoulder is, or where a particular trifocal dips. That needs the study data.

Section 07 — What the Pooled Trials Say

Intermediate is where
EDOF leads. Near is
where it trails.

Individual curves are hard to compare, so the evidence is easier to read in pooled analyses. The pattern across several systematic reviews is consistent, though the certainty varies.

AnalysisWhat it foundCaveat stated by the analysis
Karam et al., 2023 (systematic review and meta-analysis)EDOF lenses performed better at intermediate distance, and trifocal lenses better at near, on defocus curves; halos, glare and contrast did not differ significantly in the pooled analysisHeterogeneous lenses and chart methods; the near-acuity gap favours trifocal
Tavassoli et al., 2024 (Cochrane review)Distance acuity similar for trifocal and EDOF; near acuity favoured trifocalCertainty rated low; only one EDOF lens included; follow-up 3 to 6 months
Cho et al., 2022 (Bayesian network meta-analysis)EDOF gave better uncorrected intermediate acuity than monofocal, with no significant difference in contrast sensitivity, glare or halosNetwork inference across mixed lens platforms
How to use this. A wide intermediate range is a real, documented advantage of EDOF designs over monofocals. It is not the same as spectacle independence, and it does not close the near gap that the reviews describe.

Section 08 — The Agaaz Range

An EDOF lens in our
range, read the same way.

Agaaz supplies an extended depth of focus intraocular lens, X-VIZ EDOF. The same reading discipline applies to it as to any other lens on this page.

X-VIZ EDOF — extended depth of focus IOL
Intraocular lens
The product page describes its own range-of-vision curve. As that page states, the curve is a relative measure without an absolute scale, so it cannot be read on a logMAR axis or compared directly with the curves in the studies above. Ask for the chart, threshold, pupil size, number of eyes and monocular or binocular testing for any curve shown, including this one.

For the full specification, see the X-VIZ EDOF product page. For the wider lens range, see the complete portfolio.

Section 09 — FAQ

Frequently asked questions
about defocus curves.

A defocus curve plots visual acuity against defocus, the amount of focus shifted away from distance, usually measured in dioptres. It shows how well the lens keeps vision sharp across a range of focus rather than at a single point. The curve is only meaningful once you know the chart used to measure acuity, the threshold applied, the pupil size, the number of eyes tested and whether testing was monocular or binocular.

In the scoping review by Fernández and colleagues (2023), an EDOF classification is set against a monofocal reference using two tests: a gain of at least 0.5 D in depth of field at 0.2 logMAR, and intermediate visual acuity of 0.2 logMAR or better. The threshold is a convention, so the same curve can pass or fail depending on how it was measured.

Three things change the shape. The chart matters: different charts can give different acuity at the same defocus. The curve fit matters: different mathematical fits to the same data can over- or under-state the curve. The defocus steps and range matter. Curves from different studies are therefore not directly comparable unless these methods match.

Not by itself. A wider curve describes how much range of focus is kept above a chosen threshold, mainly at intermediate distances. It says little about near reading, where the pooled trials generally favour trifocal designs, and it does not measure how often patients still need spectacles in daily life.

Yes. Bench studies of extended-depth-of-focus lenses have shown that the through-focus pattern depends on pupil diameter and on the eye's spherical aberration. At some pupil sizes the foci merge into one broad zone; at others the pattern can look bifocal-like. A curve measured at one pupil size does not describe behaviour at another.

References & Evidence Base

Peer-reviewed
citations.

Buckhurst PJ, Wolffsohn JS, Naroo SA, et al. "Multifocal intraocular lens differentiation using defocus curves." Invest Ophthalmol Vis Sci. 2012. PMID 22589435
Law EM, Buckhurst HD, Aggarwal RK, et al. "Optimising curve fitting techniques to look for standardisation of the analysis of defocus curves derived from multifocal intraocular lenses." Ophthalmic Physiol Opt. 2022. PMID 35403738
Tañá-Rivero P, Tañá-Sanz P, Tañá-Sanz S, et al. "Recognition vs resolution charts for defocus curve determination in trifocal intraocular lenses." J Cataract Refract Surg. 2024. PMID 38780434
Fernández J, Rocha-de-Lossada C, Zamorano-Martín F, et al. "Positioning of enhanced monofocal intraocular lenses between conventional monofocal and extended depth of focus lenses: a scoping review." BMC Ophthalmol. 2023. PMID 36918799
Chae SH, Son HS, Khoramnia R, et al. "Laboratory evaluation of the optical properties of two extended-depth-of-focus intraocular lenses." BMC Ophthalmol. 2020. PMID 32059666
Kozhaya K, Kenny PI, Esfandiari S, et al. "Effect of spherical aberration on visual acuity and depth of focus in pseudophakic eyes." J Cataract Refract Surg. 2024. PMID 37702444
Tabernero J, Otero C, Kidd J, et al. "Depth of focus as a function of spherical aberration using adaptive optics in pseudophakic subjects." J Cataract Refract Surg. 2025. PMID 39848274
Karam M, Alkhowaiter N, Alkhabbaz A, et al. "Extended Depth of Focus Versus Trifocal for Intraocular Lens Implantation: An Updated Systematic Review and Meta-Analysis." Am J Ophthalmol. 2023. PMID 36736751
Tavassoli S, Ziaei H, Yadegarfar ME, et al. "Trifocal versus extended depth of focus (EDOF) intraocular lenses after cataract extraction." Cochrane Database Syst Rev. 2024. PMID 38984608
Cho JY, Won YK, Park J, et al. "Visual Outcomes and Optical Quality of Accommodative, Multifocal, Extended Depth-of-Focus, and Monofocal Intraocular Lenses in Presbyopia-Correcting Cataract Surgery: A Systematic Review and Bayesian Network Meta-analysis." JAMA Ophthalmol. 2022. PMID 36136323

Continue Reading

Related pages
from Agaaz.