From Keratome to Capsular Bag: How IOL Injectors, Cartridges & Viscoelastic Change the Wound (2026 Guide)
The nominal incision is not the final incision: what laboratory, animal and clinical studies measured for wound enlargement, injector tip size and viscoelastic backflow.
From Keratome to Capsular Bag: How IOL Injectors, Cartridges & Viscoelastic Change the Wound | Agaaz Ophthalmics

Surgical Delivery · Surgeon Series

The nominal incision
is not the
final incision.

The keratome sets a width, but the wound keeps changing: at phacoemulsification, then again when the injector tip and the lens pass through. This guide sets out what laboratory, animal and clinical studies measured at each step, what drives the enlargement and where the evidence stops.

0.1–0.65 mmwound enlargement range
across 13 injector models
(rabbit model)
5 of 13systems needed a larger
incision than recommended
(porcine, ex vivo)
0 trialshere measure infection
for preloaded vs
manual loading

Section 01 — The Wound Grows at Every Step

A keratome width is a starting
point, not a result.

Cataract surgeons choose a blade by width and an injector by its recommended incision. Both numbers describe where the wound begins. Measurements taken inside the eye show where it ends up.

Steinert and Deacon measured 46 eyes with an internal incision gauge at three points: after the initial keratome entry, after cataract removal, and after a folded lens was implanted. The incision made by a diamond keratome was wider than the keratome's physical width by a mean of 0.16 mm. After phacoemulsification and irrigation/aspiration it widened by a further mean of 0.09 mm. Implantation of the folded silicone lens, by forceps or by injector, widened it by a mean of 0.26 mm more. Widening the incision before insertion did not remove the extra expansion. The authors concluded that the incision stretches or tears irreversibly at each step, and that studies assuming the initial keratome size equals the final size may be wrong.

That series used silicone lenses and devices from the 1990s, so its numbers are not a description of today's injectors. Mamalis later measured 100 eyes and found the same direction of effect: clear corneal incisions enlarged in a predictable way after foldable lens insertion, and lenses delivered by forceps enlarged the wound by more than lenses delivered by injector (about 3% for injected lenses, against roughly 4 to 6% with forceps).

Step 1: keratome entry
The incision measured wider than the keratome's physical width in the 1996 series (mean 0.16 mm).
Step 2: phaco and aspiration
Instruments passing through the tunnel widened it further (mean 0.09 mm in that series).
Step 3: lens delivery
Forceps or injector insertion added the largest single step in that series (mean 0.26 mm), and it is the step modern injector studies focus on.

Section 02 — Interactive Figure

Step through the growing
wound, one stage at a time.

Choose a stage to see the wound drawn wider at each step. The quoted means come from the 1996 measurement series described above. The drawing exaggerates the change so it can be seen.

The wound at each surgical step (surface view)
Data: the means in the readout are from Steinert and Deacon, 1996 (46 eyes, silicone lenses). Schematic: the drawn width is exaggerated and not to scale, and it does not predict the final size of any modern injector.
Grey ticks: the nominal width. Gold bracket: the wound after the selected step. Distances in the drawing are not to scale.

Section 03 — How Much With Modern Injectors?

Ranges, not a single
number for every injector.

Four experimental studies measured wound enlargement with modern injector systems. They used different models, different devices and different starting incisions, which is why the table reports each one separately.

StudyModelWhat it reportedLimit stated or implied
Arboleda et al., 2019Rabbit surgery; 71 injectors of 13 modelsInjector tip diameters ranged from 1.44 to 2.12 mm. Initial incisions ranged from 2.0 to 2.63 mm. All injectors enlarged the wound, by 0.1 to 0.65 mm, and no final incision measured under 2.0 mm.Animal model; a co-author is affiliated with an injector manufacturer
Friedrich et al., 2023Porcine eyes ex vivo; 13 injector systems; 499 incisions87% of incisions enlarged, by a mean of 0.26 ± 0.18 mm. Five systems needed a final incision about 0.3 mm larger than the manufacturer's recommendation.Ex vivo animal tissue; the senior author discloses industry fees and grants
Oshika & Wolfe, 2019Porcine eyes; 4 preloaded systems; 10 eyes per groupA tighter incision and cartridge-insertion technique caused more enlargement. A smaller starting incision did not necessarily give a smaller final incision.Small groups; a co-author is employed by a manufacturer
Oshika & Sasaki, 2021Porcine eyes; one automated preloaded system; 10 eyes per groupThat system caused less enlargement than the other preloaded systems it was compared with. With a 2.4 mm corneal incision the enlargement was smaller than with 2.2 mm, but the final incision was still smaller with 2.2 mm.One device; comparison drawn against an earlier study; a co-author is employed by a manufacturer
How to read the table. Every study found enlargement, and every one found that it varied by system. The size of the effect depends on the model used to measure it, so a figure from one injector or one tissue cannot be transferred to another. Animal and ex vivo findings also have not been shown to predict clinical outcomes by themselves.

Section 04 — What Drives the Final Size

Starting incision, tip diameter,
depth and technique.

If the wound grows, the next question is which choices change by how much. One laboratory study analysed this directly.

Friedrich and colleagues analysed 409 clear corneal incisions made with 126 injectors of 13 models in 126 cadaver porcine eyes, then used a mixed-effects model to rank the influences. A larger final incision was associated with a larger starting incision, a larger vertical diameter of the injector tip and a deeper insertion, while a longer incision tunnel was associated with a smaller final incision. The starting incision had the largest effect, followed by the tip's vertical diameter, insertion depth and incision length. Corneal thickness and the injector's tip angle did not have a significant effect. The authors state that further studies are needed to confirm the findings in human autopsy eyes and in clinical practice.

Starting incision
The largest single association with final size in that model. A tighter start also gave more enlargement in the Oshika study, which is why a smaller label does not always mean a smaller final wound.
Tip vertical diameter
The second strongest association. The authors conclude that the injector's vertical diameter should be as small as possible.
Insertion depth and tunnel length
Shallower insertion and a longer tunnel were associated with a smaller final size. Both are technique choices, not device properties.
The delivery force side
A narrower nozzle also changes what the lens goes through

Cabeza-Gil and colleagues measured the force needed to eject seven lenses through four syringe-type injectors of different sizes. Lens material and cartridge nozzle size were the most influential parameters. Every lens could be delivered through a 2.2 mm injector, while the 1.8 mm injector damaged several lenses. This is a bench study of syringe-type injectors, so it does not describe preloaded systems directly, but it shows that nozzle size is a constraint on the lens as well as on the wound.

Section 05 — The Clinical Comparisons

Real eyes, small differences,
similar outcomes.

Three clinical studies compared two preloaded systems in patients. They are the closest this evidence set gets to the question surgeons ask, and they all compared one preloaded system with another.

StudyDesignWhat it reportedLimit stated or implied
Yildirim et al., 2021Prospective, randomised, intra-individual; 58 paired eyesEnlargement was 0.20 ± 0.10 mm for one system and 0.29 ± 0.10 mm for the other, a significant difference. Final wound sizes were 2.41 and 2.35 mm. Astigmatism and 3-month visual acuity were similar.Two systems; the starting incisions differed (2.2 and 2 mm); 3-month follow-up
Khoramnia et al., 2023Prospective, randomised, intra-individual; 96 eyes of 48 patients, 2.2 mm startOne system enlarged the wound less than the other when its insert shield was used (0.213 vs 0.265 mm). Without the shield the difference was not significant (0.272 vs 0.296 mm). Astigmatism and acuity were comparable, so the authors judged the difference to be of limited clinical relevance.Two systems; 3-month follow-up
Haldipurkar et al., 2020Prospective, non-randomised; 276 eyes of 220 patients, 2.2 mm startMean incision size after lens insertion was 2.367 mm and 2.369 mm, with no significant difference in incision change, acuity or astigmatism.Non-randomised; two systems; one institution
What this means. Differences in enlargement between systems were real but small, on the order of a few hundredths of a millimetre to about a tenth. In these studies they did not translate into different astigmatism or acuity at the time points measured. That is reassuring for the systems tested, and it says nothing about systems that were not tested.

Section 06 — The Viscoelastic Link

What happens to the viscoelastic
at the cartridge tip?

Viscoelastic is used in the anterior chamber and in the cartridge. One bench study looked at which way it flows when the lens is delivered.

Matsuura and Inoue loaded each injector with coloured viscoelastic at the cartridge tip and observed where the viscoelastic went as the lens was inserted, using three injector and lens combinations, five of each. In the 2.5 mm combination with a three-piece lens, intraocular viscoelastic flowed back into the injector and pushed the coloured viscoelastic deep into the cartridge without it entering the eye. In the 2.7 mm preloaded combination, a large amount of intraocular viscoelastic also flowed back and only a small amount of coloured viscoelastic entered the eye. In the 2.2 mm preloaded single-piece combination, backflow was limited and a large amount of coloured viscoelastic entered the eye with the lens.

The authors' reading is that backflow into the cartridge can limit intraocular contamination, but that small-diameter cartridges with plate-type haptics allow insufficient backflow. They also note the surgical field cannot be made completely sterile. This is a mechanism study of a handful of devices. It did not count infections, and nothing in it shows that any injector or viscoelastic raises or lowers the rate of endophthalmitis.

The practical point. A narrower cartridge changes more than wound size. Where the viscoelastic goes is part of how a delivery system behaves, and it is worth asking a supplier what was tested, without assuming a result that the study did not measure. The effect of viscoelastic choice on this flow was not tested in the study. See our guides to cohesive and dispersive viscoelastics and to pressure spikes after surgery.

Section 07 — The Agaaz Range

Preloaded delivery in our range,
described by specification.

Agaaz supplies the X-VIZ extended depth of focus lens in preloaded form. This section describes what the product page states and what it does not, in the same terms as the studies above.

X-VIZ — preloaded delivery systems
Extended depth of focus IOL
The product page describes the lens as supplied in two preloaded systems: an Agaaz standard preloaded system and a system from Medicel, imported from Switzerland, with a step-by-step method of use for each. It lists incision sizes of 2.6, 2.8 and 3.0 mm for the standard system and 2.4, 2.6 and 2.8 mm for the Medicel system. The page does not publish a measured final incision size or a wound-enlargement figure for either system, and none of the studies in this guide tested them.
How to use this. The listed incision sizes are recommendations for use. They are not measured final wound sizes, and the evidence above says the two can differ. If you are comparing delivery systems for a unit, including ours, ask for the measured final incision at the stated starting incision, how it was measured and on what model.

For the blades that make the starting incision, see our guide to surgical blade types and our companion piece on what trials measure for incision size and wound seal. The full range is listed in our product portfolio.

Section 08 — Checklist & FAQ

Questions to ask before
you compare injectors.

Before you accept an incision claim
Ask the study, and the supplier, these

Was the final incision measured after delivery, or is the figure the recommended starting size? In what model (human, rabbit, porcine ex vivo), and with how many eyes? What was the nozzle's outer diameter and the lens material? Was the comparison between devices on the same starting incision? Who funded it and who wrote it? Did the study measure outcomes in patients, or only the wound?

Not necessarily. Measurements show the wound enlarges as the injector tip and lens pass through it. In a laboratory study of 13 injector systems in porcine eyes, most incisions enlarged, and five systems needed a final incision about 0.3 mm larger than the manufacturer's recommendation. The label is a starting point, and the measured final size is the number to ask for.

It depends on the system, the starting incision and the technique. A rabbit-model study of 13 injector models reported enlargement of 0.1 to 0.65 mm, with no final incision under 2.0 mm. A porcine ex vivo study reported a mean enlargement of about 0.26 mm. In two clinical comparisons of preloaded injectors, mean enlargement was roughly 0.2 to 0.3 mm. These are different models and different devices, so the figures are ranges, not a promise.

The clinical comparisons reviewed here compared one preloaded system with another, not preloaded with manual loading, so they cannot answer that question. They found small differences in wound enlargement between systems, and the final incision size, astigmatism and visual acuity were similar or not significantly different. No study here measured infection rates.

A bench study using coloured viscoelastic found that backflow of intraocular viscoelastic into the injector occurred and varied between injector and lens combinations. With the smallest-incision (2.2 mm) preloaded combination, backflow was limited and a large amount of the coloured viscoelastic at the tip entered the eye. The study describes a mechanism for possible contamination, and it did not measure infection rates.

Ask for the incision size used in testing, the measured final incision size after delivery, how it was measured and on what model, the injector's nozzle outer diameter, and which lens materials and powers were tested. A recommended incision size without a measured final size leaves out the part that changes at the wound.

References & Evidence Base

Peer-reviewed
citations.

Steinert RF, Deacon J. "Enlargement of incision width during phacoemulsification and folded intraocular lens implant surgery." Ophthalmology. 1996. PMID 8594505
Mamalis N. "Incision width after phacoemulsification with foldable intraocular lens implantation." J Cataract Refract Surg. 2000. PMID 10683791
Arboleda A, Arrieta E, Aguilar MC, et al. "Variations in intraocular lens injector dimensions and corneal incision architecture after cataract surgery." J Cataract Refract Surg. 2019. PMID 30857855
Friedrich M, Auffarth GU, Merz PR. "Experimental analysis of recommended corneal incision sizes in cataract surgery using 13 intraocular lens injector systems." Sci Rep. 2023. PMID 36792674
Friedrich M, Baur ID, Yildirim TM, et al. "Laboratory Analysis of Causative Factors for the Final Incision Size due to Intraocular Lens Injector Insertion." Ophthalmol Sci. 2024. PMID 37869017
Oshika T, Wolfe P. "In vitro comparison of delivery performance of 4 preloaded intraocular lens injector systems for corneal and sclerocorneal incisions." J Cataract Refract Surg. 2019. PMID 30853318
Oshika T, Sasaki N. "Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions." J Ophthalmol. 2021. PMID 33859834
Yildirim TM, Łabuz G, Baur ID, et al. "Corneal Incision Enlargement in Two Preloaded Intraocular Lens Injectors: An Intraindividual In Vivo Study." J Refract Surg. 2021. PMID 34044690
Khoramnia R, Baur ID, Łabuz G, et al. "Enlargement of main corneal incision: clinical intraindividual comparison of two preloaded intraocular lens injectors." J Cataract Refract Surg. 2023. PMID 36209736
Haldipurkar SS, Shetty V, Haldipurkar T, et al. "Incision size changes after cataract surgery with intraocular lens implantation: comparison of 2 preloaded IOL implantation injectors." J Cataract Refract Surg. 2020. PMID 32126035
Cabeza-Gil I, Ríos-Ruiz I, Calvo B. "Experimental evaluation of the injection force exerted in intraocular lens delivery with syringe-type injectors." J Mech Behav Biomed Mater. 2021. PMID 34492403
Matsuura K, Inoue Y. "Ophthalmic viscosurgical device backflow into cartridge during intraocular lens insertion using injectors." Clin Ophthalmol. 2014. PMID 24520189